Extremes in hematocrit may affect test results. Hematocrit levels
less than 30% may cause falsely high readings. Hematocrit levels
greater than 54% may cause falsely low readings.
Inaccurate results may occur in severely hypotensive individuals
or patients in shock. Inaccurate low results may occur for individuals
experiencing a hyperglycemic-hyperosmolar state, with or without
ketosis. Critically ill patients should not be tested with blood glucose
meters.
Triglycerides up to 3,000 mg/dL do not signicantly affect test results.
However, glucose values in specimens beyond this triglycerides level
should be interpreted with caution.
Normal endogenous (within body) natural levels of uric acid,
ascorbate (vitamin C), bilirubin, triglycerides, and hemoglobin do not
interfere with your blood glucose results obtained.
Externally taken drugs L-dopa, dopamine, methyl-dopa,
acetaminophen, and ibuprofen will not interfere with
ReliOn
®
Conrm blood glucose results when taken at therapeutic
concentrations.